Docsity
Docsity

Prepare for your exams
Prepare for your exams

Study with the several resources on Docsity


Earn points to download
Earn points to download

Earn points by helping other students or get them with a premium plan


Guidelines and tips
Guidelines and tips

Equipment design for radiation protection, Lab Reports of Radiology

Equipment design for radiation protection

Typology: Lab Reports

2021/2022

Uploaded on 03/10/2023

Omsimmm
Omsimmm 🇵🇭

46 documents

1 / 65

Toggle sidebar

This page cannot be seen from the preview

Don't miss anything!

bg1
Equipment Design for
Radiation Protection
by;: Allen G. Espanola
pf3
pf4
pf5
pf8
pf9
pfa
pfd
pfe
pff
pf12
pf13
pf14
pf15
pf16
pf17
pf18
pf19
pf1a
pf1b
pf1c
pf1d
pf1e
pf1f
pf20
pf21
pf22
pf23
pf24
pf25
pf26
pf27
pf28
pf29
pf2a
pf2b
pf2c
pf2d
pf2e
pf2f
pf30
pf31
pf32
pf33
pf34
pf35
pf36
pf37
pf38
pf39
pf3a
pf3b
pf3c
pf3d
pf3e
pf3f
pf40
pf41

Partial preview of the text

Download Equipment design for radiation protection and more Lab Reports Radiology in PDF only on Docsity!

Equipment Design for

Radiation Protection

by;: Allen G. Espanola

RADIATION SAFETY FEATURES OF RADIOGRAPHIC

EQUIPMENT, DEVICES, AND ACCESSORIES

  • (^) Protective Tube Housing- A lead-lined metal protective tube housing required to protect the patient and imaging personnel from off-focus, or leakage radiation by restricting the emission of x-rays to the area of the useful, or primary, beam
  • (^) must be constructed so that the leakage radiation measured at a distance of 1 m from the x-ray source does not exceed 1 mGya/hr (100 mR/hr) when the tube is operated at its highest voltage at the highest current that allows continuous operation

Control Panel, or Console

  • (^) where technical exposure factors such as milliamperes (mA) and peak kilovoltage (kVp) are selected and visually displayed
  • (^) must be located behind a suitable protective barrier that has a radiation-absorbent window that permits observation of the patient during any procedure
  • (^) must indicate the conditions of exposure and provide a positive indication when the x-ray tube is energized.
  • (^) The visible mA and kVp digital readouts permit the operator to assess exposure conditions

Radiographic Examination Table

  • (^) must be strong and must adequately support the patient
  • (^) Frequently, this piece of equipment has a floating tabletop that makes it easier to maneuver the patient during an imaging procedure
  • (^) The thickness of the tabletop must be uniform, and for undertable x- ray tubes as used in fluoroscopy, the patient support surface also should be as radiolucent as possible so that it will absorb only a minimal amount of radiation, thereby reducing the patient's radiation dose
  • (^) A carbon fiber material is commonly used in the tabletop to meet this requirement

X-Ray Beam Limitation Devices

  • (^) Collimators
  • (^) Aperture diaphragms- simplest , used in trauma radiographic units
  • (^) Cones- replaced by collimators, used in coned view of the paranasal sinuses and sella turcica
  • (^) Cylinders – same as cones but different in shape, flared down or cylindrical
  • (^) these devices confine the useful, or primary, beam before it enters the area of clinical interest and thereby limit the quantity of body tissue irradiated.
  • (^) This also reduces the amount of scattered radiation in the tissue and prevents unnecessary exposure to tissues not under examination
  • (^) the patient's dose is significantly reduced because a smaller field size produces less scatter radiation. This improves the overall quality of the radiographic image
  • (^) the collimator is the most versatile device for defining the size and shape of the radiographic beam
  • (^) The light-localizing variable-aperture rectangular collimator is the type of collimator most often used with multipurpose x-ray units
  • (^) Two sets of adjustable lead shutters mounted within the device at different levels
  • (^) A light source to illuminate the x-ray field and permit it to be centered over the area of clinical interest
  • (^) A mirror to deflect the light beam toward the patient to be radiographed

Beam Alignment

  • (^) It is imperative that the x-ray beam and the image receptor be correctly aligned with each other
  • (^) Every radiographic tube must have a device in place to ensure accurate beam alignment

Types of Filtration

  • (^) Inherent filtration- Glass envelope encasing the x-ray tube ,Insulating oil surrounding the tube ,Glass window in the tube housing
  • (^) Added filtration- Sheets of aluminum (or the equivalent) of appropriate thickness
  • (^) Total filtration of 2.5 mm aluminum equivalent for fixed x-ray units operating above 70 kVp is the regulatory standard
  • (^) radiographic equipment requires total filtration of 1.5 mm aluminum equivalent for x-ray units operating at 50 to 70 kVp, whereas fixed units operating at below 50 kVp require only 0.5 mm aluminum equivalent
  • (^) Mobile diagnostic units and fluoroscopic equipment require a minimum of 2.5 mm aluminum equivalent.

Filtration for Mammographic Equipment