








Study with the several resources on Docsity
Earn points by helping other students or get them with a premium plan
Prepare for your exams
Study with the several resources on Docsity
Earn points to download
Earn points by helping other students or get them with a premium plan
Community
Ask the community for help and clear up your study doubts
Discover the best universities in your country according to Docsity users
Free resources
Download our free guides on studying techniques, anxiety management strategies, and thesis advice from Docsity tutors
You will learn from this practical to determine the concentration of a colored food dye, Erythrosin B, in an unknown solution
Typology: Lab Reports
1 / 14
This page cannot be seen from the preview
Don't miss anything!
Revised 10/24/
Measurement of Absorbance A spectrophotometer (Spec 20) will be used in this experiment to determine the absorbance of light absorbed by the colored Erythrosin B solution (Figure 1). Figure 1 Spec 20 Beer’s Law and the Determination of Concentration of an Unknown In order to find the concentration of an unknown, you will need to make a calibration curve using a set of solutions with known concentrations. Beer’s law shows the relationship between absorbance and concentration as follows: A = e l C where A = Absorbance (no unit) e = molar absorptivity (M-^1 cm-^1 ) l = path length (1.00 cm in this experiment) C = concentration (in mol/L) Some books refer to Beer’s law as just A = b C where “b” combines both the molar absorptivity and path length. When plotting absorbance versus concentration, a calibration curve or standard absorbance curve is established in the form of a straight line since the molar absorptivity (e) is constant for a given substance and the path length ( l ) of the cuvette is constant. The (0,0) point is included in the data when generating the calibration curve as the solvent was used to calibrate the Spec 20 at 100% transmittance which is equivalent to 0 absorbance. Therefore, for 0 concentration of solute zero absorbance is measured. The equation of the line will then be used to calculate the concentration of an unknown upon measuring the absorbance of the unknown. Wavelength filter lever
Figure 2 Example of an absorbance curve with the trend-line forced through zero The slope of the line in Figure 2 is 70, 193 M-^1 cm-^1 and the molar absorptivity times the pathlength is 70,193 M-^1 since the path length is 1.00 cm. The calibration curve shows good linearity as R^2 is very close to 1 (R^2 =0.99556). Figure 2 is an EXAMPLE not necessarily the concentration values needed to generate the curve in this experiment. The calibration curve in Figure 2 shows 6 data points, one at (0,0) and the others uses different solutions of known concentration. Although it is better to have more points on a calibration curve than just two, one must consider time required to do the experiment. Therefore, think of what would be the minimum number of points to generate a decent calibration curve. Also, the absorbance of your unknown must fall within the range of your calibration points. Limit to Linearity of the Spec 20 One must also be aware that instruments have a limit to their linearity. For example, it is common for absorbance measurements to “level off” and no longer remain linear beyond a certain concentration of solution. For the red dye solution, this usually occurs around 0.8 or higher absorbance. For example, in Figure 2, the absorbance starts to level off and the relationship between absorbance and concentration no longer is linear beyond 0.8 absorbance. Figure 3 shows what would happen if one were to continue making more solutions of higher concentration beyond 0.8 absorbance for a red dye solution.
0 2 4 6 8 10 12 Absorbance Concentration (x10-^3 M) y = 70,193 x R^2 = 0.
Figure 3 Calibration curve no longer linear beyond an absorbance of 0.8. When developing your procedure in your lab notebook, it is important to be considerate of a few things. Keep in mind that your procedure has to be detailed enough for someone else to duplicate your experiment even if they have never seen this experiment before. In addition, careful thought must be exercised in your procedure to consider limitations on precision of glassware as this ultimately affects the precision of the standard calibration curve, and hence, your determination of the concentration of your unknown. In addition, the accuracy will be affected by the quality of lab skills and tools utilized. When scientists look at others’ published experimental work, they think about using select procedures and lab glassware that will yield the best precision and accuracy to yield quality results so that the results from the experiment can be trusted. In addition, one might first measure the absorbance of the stock solution used in this experiment to see if it is even near 0.8 absorbance. If it is much higher, a dilution may be required to get the stock solution within range of the linearity of the Spec 20. This can be a difficult task. Therefore, a hint to doing this is to make a solution that is 5 or 10 times less concentrated than the stock solution. Your grade for this experiment may depend on your notebook recording and quality of results (precision and accuracy). Your professor may also require a typed lab report in part or in full. Ask your professor for specific grading criteria.
0 5 10 15 20 25 Absorbance Concentration (x10-^3 M)
Erythrosin B (Acid Red 3) Stock Solution (molarity approximately 8 .00x10-^5 M) EQUIPMENT TO BE CHECKED OUT 100 mL, 50 mL, 25 mL, and/or 10 mL volumetric flasks can be used (only one type of volumetric flask is allowed per student) 2, 5, 10 mL graduated pipets can be used Pipet bulb or pipet pump Cuvets for the spectrophotometer SAFETY Goggles must be worn at all times in the bench and working areas of the laboratory whenever anyone is working. If goggles become clouded, go to the non-bench area of the lab, remove goggles, and clear them. Goggles may be removed in the seating area of the laboratory. Keep solutions away from skin or face, and wash off immediately if chemicals touch the skin. WASTE DISPOSAL All solutions may be poured down the drain EXPERIMENTAL PROCEDURE: This experiment is designed for students to work independently from one another. Do not consult with students during this lab experiment. Be aware that you may only have two lab periods to make a calibration curve and determine the concentration of an unknown. Therefore, you will need to be efficient and work quickly. Rinse any glassware that was used with the dye solution immediately when you are finished using it as the dye can be difficult to clean from glassware. Part 1: Calibrate the Spec 20 and Check absorbance of a diluted red dye solution Write a procedure for part 1 for calibrating the Spec 20, making a dilution and measuring the absorbance of the diluted red dye solution. Use the following to help you write your procedure:
Written by Jody Williams Tyler with helpful contributions from discussions with Tatiana Lopez, Kamran Golestaneh, and Terri Beam and review comments by Terri Beam and Kamran Golestaneh from the Chemistry Department at Mt. San Antonio College. Stock, J.L., J. Chem. Ed ., 1995, 72 (10), p. 926. Beam, T., “Copper Concentration”, Chemistry 50 Laboratory Manual, Mt. San Antonio College, Fall 2012, p. 129.