Docsity
Docsity

Prepare for your exams
Prepare for your exams

Study with the several resources on Docsity


Earn points to download
Earn points to download

Earn points by helping other students or get them with a premium plan


Guidelines and tips
Guidelines and tips

Redesigning Carotid Angioplasty Filters: Current & Proposed Filters with Smaller Pore Size, Lab Reports of Biology

The research on current distal protection filters used during carotid angioplasty, specifically the filterwire ex, accunet, and angioguard. The document also presents the goal of redesigning the distal protection filter to improve its efficiency in capturing embolic particles by decreasing pore size and adding a skirt. The design of the redesigned filter, the modifications made to the flow loop for testing, and the methods used to analyze the data. The ultimate goal is to determine if the redesigned filter is more efficient than other filters and to market it as the top distal protection device.

Typology: Lab Reports

Pre 2010

Uploaded on 09/09/2009

koofers-user-x4m
koofers-user-x4m 🇺🇸

10 documents

1 / 16

Toggle sidebar

This page cannot be seen from the preview

Don't miss anything!

bg1
Redesign of a Distal Protection Filter
for Carotid Artery Angioplasty
By:
Sandeep Devabhakthuni
Chenara Johnson
Daphne Kontos
Perry Tiberio
Professor Mark Garter
Bioeng 1160 Bioengineering Design
26 April 2004
1
pf3
pf4
pf5
pf8
pf9
pfa
pfd
pfe
pff

Partial preview of the text

Download Redesigning Carotid Angioplasty Filters: Current & Proposed Filters with Smaller Pore Size and more Lab Reports Biology in PDF only on Docsity!

Redesign of a Distal Protection Filter

for Carotid Artery Angioplasty

By: Sandeep Devabhakthuni Chenara Johnson Daphne Kontos Perry Tiberio Professor Mark Garter Bioeng 1160 Bioengineering Design 26 April 2004

Introduction

The first percutaneous carotid angioplasty was performed in 1980, and since then, there has been a continuous influx of new interventional technologies^1. The main safety issue arising from this procedure is the risk of cerebral embolization. This occurs when the angioplasty balloon is inflated to clear the stenosis, resulting in the release of debris into the blood stream. This debris (emboli) travels to the brain, potentially causing a stroke. Hence a new technology was developed to increase the safety of the procedure, the neuroprotection device. Neuroprotection devices are designed to capture and remove emboli released during carotid angioplasty. There are two main types: balloon occlusion devices and distal protection filters (DPF). Since the balloon occlusion inhibits blood flow during the procedure and does not allow for continuous angiographic feedback, the distal protection filter is favored. The filters permit continuous blood flow during the procedure while still providing distal cerebral protection. The main weaknesses of the filters, however, are their pore size. Since the blood is composed of particles ranging from 2 to 120 μm^2 , determining the most efficient pore size is critical to the design of a distal protection filter. Currently, the smallest pore diameter in filter devices is about 40 μm, with the majority of filters having an 80 to 100 μm pore diameter. The largest pore diameter is about 190 μm^3. In addition to pore size, an inadequate fit to the artery allows emboli to pass the filter. The three filters that minimize these complications are the FilterWire EX (Boston Scientific, Natick, MA, USA), the AccuNet (Guidant Corporation, Indianapolis, IN) and the AngioGuard (Cordis Corporation, Miami Lakes, FL). These filters are considered to be the three best distal protection filters. The FilterWire EX (see Fig. 1) polyurethane filter has one of the largest filter volumes (15 mm long), allowing it to capture more embolic particles. The 110μm-pore filter bag is

volume, unlike FilterWire EX and AccuNet. The 100μm-pore filter bag is deployed perpendicular to the vessel. Because of the filter’s shallow basket, embolic particles overfill the basket and may escape into the vessel lumen during retrieval. Figure 2: The AngioGuard (Cordis Corporation, Miami Lakes, FL)^3 Based on the research on these current filters, our goal is to redesign the distal embolic protection filter device to be more efficient in capturing embolic particles by decreasing the pore size and adding a skirt to prevent passage of emboli. Furthermore, we propose that a nylon/nitinol filter material be used in our design to maintain the mechanical properties, while retaining viability of these two materials. This will be tested through comparison of results of in- vitro studies that mimic in vivo conditions under which the device will be operated. When the experiment is performed and the analytical results are calculated, we hope to determine that the redesign of the distal embolic protection filter is providing protection at a greater efficiency than other filters, is no more difficult to use than other filters, will be used in future studies, and be marketed as the top distal protection device.

Methods

Through analyzing previous designs, we determined that the optimal pore size should be between 70 and 80 μm. As shown in Figure 3, the redesign of the distal protection filter included a radio plaque nitinol elliptical ring with the 8.51mm being the short diameter. It had a cylindrical filter bag (28mm long) mounted on the side. The nitinol ring was supported by one

nitinol strut in order to support the filter and position it at a 20° angle to the vessel. The filter had a nylon skirt (roughly 0.1 mm) extending outward from the nitinol ring. The filter will be delivered by pushing the strut to collapse the nitinol ring and deployed by releasing the strut. The ring will default to the manufactured 20° angle. Figure 3: Picture of Redesigned Distal Protection Filter This preliminary design provided all the safety precautions necessary for embolic protection during the procedure. The filter was redesigned using the SolidWorks program. The SolidWorks files were then used as a guideline for the fabrication of a testable prototype. To build the model above, the following materials were required: nylon for the basket (Sefar Holding Inc, Freibach, Thailand), nitinol, polyimide tubing and stainless steel for the frame (New England Precision Guiding Inc, Holliston, MA.). Before testing the new filter model, the flow loop used in previous studies was redesigned in order to obtain more efficient results^7. The flow loop previously used utilized continuous flow. The new flow pump induced peristaltic movements, simulating vascular smooth muscle contractions.

model blood in the flow loop was a solution consisting of glycerol and water at a ratio of 9: resulting in a viscosity of 3.5 cP in order to have a viscosity similar to blood. After the prototype was manufactured and the in vitro flow loop was completed, training was implemented for the use of the delivery of the filter and use of the flow loop. If improperly used, the results would have been poor due to human error. In addition, the placement of the filter was crucial in gaining precise results. Therefore, five trial runs with the filter in the flow loop were performed before the actual experiment is conducted. After proper training of filter use, the new filter design was tested and results were compared to previously tested filters to determine if the new design was more efficient in capturing the embolic particles and if it was a competitive new product. Ten subsequent runs, with only 5 successful trial runs, were used for analysis. Collection of the embolic microspheres occurred at one point within the filter during each trial. While the filter was deployed within the vessel walls, the in-line filter collected the missed particles to determine efficiency while deployed. The in-line filter had a pore size of less than 100μm.

Figure 5. Redesigned flow loop schematic The flow loop is detailed in Figure 5. Per trial, 28 liters of liquid (10 L of glycerol and 18 L of water) originated in the reservoir and was pumped through the circuit back to the same reservoir. Before the flow pump was turned on, the filter was inserted through the one-way valve and deployed. Then, the peristaltic flow pump (ISMATEC, Glattbrugg, Switzerland) was set to 150 mL/min. Once the flow was stabilized, microspheres that functioned as embolic particles were injected through a one-way valve to allow for continuous flow for each trial. After the microspheres were inserted, the trial ran for five minutes before the flow pump was shut off. Once the filter was retrieved, the mass of the particles removed from the filter was weighed. This completed the use of the flow loop and the process was repeated for the next trial Once this experiment was completed, data analysis was performed. To analyze the data, the percentage device efficiency was calculated and compared with the other previously tested filters. The overall device efficiency will be determined using Equation 1. Reservoir Peristaltic Flow Pump Length = 20 cm Insertion Point of Filter (one-way valve) Point of Filter Deployment Glycerin/Water solution (9:16) Inline filter

Particles Captured by Filter

Trials Percent captured Figure 6. Graphical representation of the capture efficiency for all trials As Figure 6 shows, there were data points that were significantly low. These data points were determined to be a result of device or human error and were excluded. Out of then ten trials that were performed, five trials resulted in significantly low efficiencies due to device failure. Because the results obtained for these trials were not accurate, these results were excluded in order to determine the overall efficiency of the device more accurately. As shown in Figure 7, the modified data was plotted for the five successful trials. The plot indicates that the capture efficiency decreased as the filter use increased.

Modified Capture Efficiency

**0.

0 2 4 6 8 10 12 Trials Percent captured** Figure 7. Graphical representation using the modified data for capture efficiency Once the overall device efficiency was calculated for each of these trials, the efficiency of our filter was then compared to the efficiency of predicate devices: AngioGuard, AccuNet, and FilterWire. Data for each of these three filters collected using an old flow loop tested in previous studies was used for the comparison^7. The overall device efficiency for each of these filters was calculated and the results of these calculations are shown in Table A.2 (Appendix A). The average and standard deviation was calculated for all of the filters in order to make the comparison. Table 3 showed the comparison of the device efficiency of our filter with the average device efficiency of the other filters. As the table indicates, our filter was slightly lower in the overall device efficiency with a higher standard deviation as compared to the other filters. Table 3. Comparison of Overall Device Efficiency for the Four Filters

between the vessel lumen and the filter. This gap allowed the emboli particles to escape the filter, resulting in lower capture efficiency. The reason for these unforeseen failures was due to the human error involved in the testing and device failure. As Figure 6 indicated, there was a significant amount of variation when comparing the efficiencies for all ten trials. This significant amount was considered to be caused by excessive user error during some of the trials. Because the error was preventable, the data for these trails were not relevant in determining the overall device efficiency of the filter. As a result, the data for the five unsuccessful trials were excluded, and the device efficiency of the filter was determined by the modified data shown in Figure 7. In addition to these unexpected occurrences, there were other possible limitations that accounted for poor results. For instance, the placement of the filter was not consistent for all the trials. After each trial was performed, the flow loop had to be disassembled and the filter had to be removed in order to retrieve the captured particles for weighing. Once the filter was inserted again and the flow loop was reassembled, the filter placement was changed. Therefore, this inconsistent filter placement may have affected results. Moreover, when the filter was being retrieved, the filter was not fully collapsible. This made it easier for the particles to escape the filter. Since this retrieval mechanism was not efficient, this resulted in lower capture efficiency. With that stated, our filter also suffered from mechanical failures associated with manufacturing techniques. Consequently, these limitations generated poor results that did not accurately assess the device efficiency of our filter. Due to these limitations and user error during testing, several modifications to the design were recommended for higher capture efficiency. First, a more manageable retrieval mechanism is needed in order to minimize the loss of emboli particles when the filter is in movement. A

possible improvement would be to collapse the filter completely, which would minimize the loss of particles. Second, instead of using nylon as the material for the filter basket, using polyurethane would be more efficient because polyurethane is more elastic leading to greater flexibility. This material property would be beneficial because it can aid in the capture of the particles and prevent loss of particles as the filter is being retrieved. Finally, using nitinol tubing for the ring would allow more flexibility for the filter, which would allow the filter to collapse easily. While the current materials used are all biocompatible as demonstrated by their use in other medical devices that contact blood, using materials that are seen in the predicate devices would give our filter the necessary mechanical properties. These modifications to the design would allow the filter to be more efficient in both capturing the particles and retrieving the particles.

Appendix A: Data Collection

Table 1. Measurements of the Weight of the Particles Before and After Trials Initial Weight (mg) Filter Particle Weight (mg) Trial 1 79.6 78. Trial 2 84.7 37. Trial 3 92.7 42. Trial 4 85.4 82. Trial 5 87.3 28. Trial 6 84.7 22. Trial 7 88.0 70. Trial 8 108.4 83. Trial 9 80.7 15. Trial 10 82.0 51. Table 2. Overall Device Efficiencies for Three Filters Currently on the Market AngioGuard AccuNet FilterWire Trial 1 100 100 100 Trial 2 100 100 75. Trial 3 89.29 97.73 100 Trial 4 94.74 100 78. Trial 5 100 90.91 100