Westgard
Rules
Dr Bharat Kumar Pareek
1st year Resident
Dept. of Biochemistry
S.M.S. Medical College
pf3
pf4
pf5
pf8
pf9
pfa
pfd
pfe
pff
pf12
pf13
pf14
pf15
pf16
pf17
pf18
pf19
pf1a
pf1b
pf1c
pf1d
pf1e
pf1f
pf20

Partial preview of the text

Download WESTGARD RULE and L- J curve. and more Slides Medical Biochemistry in PDF only on Docsity!

Westgard

Rules

Dr Bharat Kumar Pareek

st

year Resident

Dept. of Biochemistry

S.M.S. Medical College

  • (^) Westgard rule AKA Multirule Quality Control was developed by Dr. James Westgard based on statistical concepts. DR JAMES WESTGARD FOUNDER

Westgard rule/Multirule Quality Control(QC)

  • (^) The formulation of Westgard rules were based on statistical methods. Westgard rules are commonly used to analyze data in LEVEY–JENNINGS (LJ) chart.
  • The advantages of multirule QC procedures are that false rejection can be kept low while at the same time maintaining high error detection.
  • Westgard rules are used to detect both random and systematic errors.

How to perform Multirule QC?  Choose a good quality control material.  (^) Run control at least 20 times (different runs) – once or twice daily  (^) Collect your control measurements  (^) Obtain the mean and Standard deviation for 20 points of data  (^) Use the data and draw Levey-Jennings(LJ) chart

Levey-Jennings (LJ) Chart

  • (^) It is a graphical method for displaying control results and helps in evaluating whether a procedure in-control or out-control by applying westgard rules.
  • (^) Daily control values are plotted versus time.
  • (^) Lines are drawn from point to point, to understand any systematic or random errors.
  • (^) Every control that we run is plotted on LJ chart. LJ chart has x and y axis. On x axis there will be no of days/time while on y axis there will be value of mean and standard deviation.

Levey-Jennings (LJ) Chart

Types of Westgard rules  Most common rules-

  1. 1 2s Rule
  2. 1 3s Rule
  3. 2 2s Rule
  4. 4 1s Rule
  5. R4s Rule
  6. 8 x Rule
  7. 10 x Rule
  8. 12 x Rule

Other rules-

  1. 2of32s Rule
  2. 3 1s Rule
  3. 6 x Rule
  4. 9 x Rule
  5. 7 T Rule 1 = how many time rule is violated 2S = which standard deviation is violated

There are 6 Basic Rules

1 3s Rule:- 1 3s control rule is commonly used with a Levey-Jennings chart. It refers when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

2 2s Rule:- 2 2s refers when 2 consecutive control measurements exceed the same mean plus 2s or the same mean minus 2s control limit. It requires corrective action and run is rejected

R4S Rule:- R 4s

  • refers when 1 control measurement in a group exceeds the mean minus 2s and another exceeds the mean plus 2s.This rule should only be interpreted within run ,not between run. Test run must be rejected.

8 x Rule:- 8x – refers when 8 consecutive control measurement fall on the one side of mean. Test run must be rejected.

12 x Rule:- 12x – refers when 12 consecutive control measurements fall on one side of the mean. Test run must be rejected.

  • (^) This decision tree helps to decide whether to reject or accept a run. If your answer is YES, it means that rule is violated.